Due to its scan geometry, the Hologic system makes one pass of the region of interest with a broad fan beam. Thus, the X-rays pass through a part of the body only once. A consequence of the Hologic
geometry is that bone area is magnified based on the distance between the examination table and the spine. In contrast, with its narrow fan beam, Prodigy scanners make multiple passes and over samples some parts of the scan area while not sampling other areas at all depending on where the passes intersect above the tabletop. The Prodigy scanner stitches the passes together in the bone plane to create an undistorted view of the bone. The Prodigy does not selleck chemicals exhibit magnification [5]. Another consideration is that the 1994 sBMD study was derived from data collected at one clinic using one system from each manufacturer and could not take into account intra-manufacturer variation. Our study consisted of three study sites, with
three pair of Hologic Delphi and GE-Lunar Prodigy devices, and the inter-site variations were intentionally not cross-calibrated to provide a more robust relationship. This is different than the quality control performed for multi-center clinical trials where the goal is to remove systematic differences between DXA systems by phantom cross-calibration. The difference in L2-L4 AREA showed a significant trend as function of mean AREA measured. Two possible explanations for this are the more pronounced magnification in the Hologic Delphi fan-beam systems than the GE-Lunar Prodigy and the difference in leg positioning. Boudoueq et al. [5] found in phantom experiments that decreasing height above selleck screening library the table increased AREA for the Hologic Discovery PD-0332991 price device and not for the Prodigy. Secondly, Hwua et al. found that the GE-Lunar Prodigy BMD results for the legs down position were on average 1.33% higher than when measured with legs up due to a change in the bone projection Edoxaban [17]. However,
Nord et al. showed that the GE-Lunar Prodigy spine AREA, BMC, and BMD in leg down position were highly correlated with results from the traditional position [18]. Unfortunately, we were not able to determine which of these effects accounted for the differences found in this study. This study had several limitations. First, no phantom cross-calibration was performed between study sites. The absolute calibration differences between the systems of the same make was not known during the period of the study. However, the sites were monitored with their local quality control phantoms and found to be stable and calibrated to their factory standards. Clinical systems can vary in their absolute calibration by as much as ±2% [14]. Using another set of systems may generate equations slightly different because of this. However, since there is no gold standard phantom for field calibration of either Hologic or GE-Lunar systems, this limitation is unavoidable.
Related posts:
- Index cases were asked to pass on study invitations to their firs
- J Bone Miner Metab 26:400–405CrossRefPubMed 33 Brownbill RA, Ili
- Every processor consisted of a high-pass filter, a first degree d
- While this makes obvious sense, there has been far too little cl
- Surgery The
anatomy of the bony pelvis makes complete tum