9 and 10 However, all of these methods have limitations such as long run times and/or expensive. The present study focused on minimizing these limitations and to develop a simple precise accurate and economic method for estimation of diazepam in tablet dosage form. Figure options Download full-size image Download as PowerPoint slide An analytically pure sample of diazepam was procured as gift sample
from Natco Pharma Ltd. (Hyderabad, India). HPLC grade methanol was procured from E. Merck (Hyderabad). Liquid chromatographic grade water was obtained by double distillation and purification through Milli-Q water purification system. Potassium dihydrogen phosphate (AR grade, purity 99.5%) was procured from Qualigens. Tablet formulations VALIUM (Nicholas Piramal India Ltd.) was procured from a local pharmacy with labeled amount 5 mg per tablet. The HPLC analysis was performed on CYBERLAB OSI 906 HPLC equipped with an LCP-100 reciprocating HPLC pump. A manually operating Rheodyne
injector with 20 μL sample loop, a LC-UV 100 ultraviolet detector was used. Chromatographic analysis was performed on a Hypersil reversed phase C-18 column with 250 × 4.6 mm i.d. and 5 μm particle size. The mobile phase consist of acetonitrile, methanol, 1% phosphate buffer (pH-3) in ratio of 18:58:24 (v/v/v) that was set at a flow rate of 1 ml/min. The mobile phase was degassed and filtered through 0.25 μm membrane filter before pumping into HPLC system. The eluent was monitored by UV detection at 232 nm. Stock solution of diazepam (1 mg/ml) selleck products was prepared by transferring 25 mg
of drug in a 25 ml volumetric flask. The drug is dissolved in sufficient amount of 0.1 N HCl Rolziracetam and finally the volume was made up to the mark with distilled water. Working standard solutions ranging from 0.5 to 50 μg/ml were prepared by appropriate dilutions of the stock with distilled water. Twenty tablets of diazepam hydrochloride were weighed and ground into a fine powder. A quantity of powder equivalent to 25 mg of diazepam was weighed and transferred into a 25 ml volumetric flask and was dissolved in 0.1 N HCl. The volume was made up to the mark with the same. Above solution was suitably diluted with distilled water. From this stock, appropriate dilution (10 μg/ml) was prepared. The solution thus prepared was filtered through 0.45 μ membrane filter and the resulting filtrate was sonicated for 10 min. After setting the chromatographic conditions and stabilizing the instrument to obtain a steady baseline, the sample solution was loaded in the 20 μl fixed – sample loop of the injection port. Initial trial experiments were conducted, with a view to select a suitable solvent system for the accurate estimation of the drug and to achieve good retention time.
Related posts:
- The crashed tobacco of each cigarette was quantitatively suspende
- NART methods physeal St changes Hypoxia and determination
- Andarine GTX-007 was measured by using the method described above for TFV
- Current diagnostic methods for ACS in the ED, however, are clearl
- In this respect, the sensitivity variation as a function of the